9.3.2 Tolerance check after setting the limit values by weighing
1.
Press the SET-key and keep it pressed until the
display for setting the lower limit value „L.SET
appears. The current setting flashes.
2.
Place the nominal weight (e.g. 95 g) for the lower
limit value and confirm with PRINT-key.
For menu setting „23.Pi. 1“ the input finishes here.
Start tolerance check, see step 4.
For menu setting „23.Pi. 2“ wait until the display
„H.SET“ for setting the upper limit value appears.
The current setting flashes.
3.
Place the nominal weight (e.g. 105 g) for the upper
limit value and confirm with PRINT-key.
Remove the nominal weight. From here evaluation
takes place whether the goods to be weighed are
within the two tolerance limits.