40 Keysight E5061B Network Analyzer
Troubleshooting
System Recovery
2-
System Recovery
By executing system recovery, you can return the E5061B’s system (the
Windows operating system and the firmware) to the factory state (at the time
of purchase
*
).
Notes on executing the factory recovery function
Executing the factory recovery function causes the following conditions:
In addition to the Windows operating system and the firmware, the
following settings of the E5061B are returned to the factory state.
a. Network setting
b. GPIB setting
c. Printer setting
The driver for the supported printer which is installed after purchase is
deleted.
You need to execute initial registration again.
Files that you have created using the save function (files in the D drive) are not
affected. However, Keysight recommends backing them up before executing
system recovery for precautionary purposes. For more information on backup,
refer to “Backing Up the Data” as described in E5061B Online Help.
Procedure to execute the factory recovery
This section describes how to return the contents of the C drive to the factory
state.
Step 1. Shut down the E5061B.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5061B.
Step 4. Press the standby switch of the E5061B to turn it on.
*. The default setting of the hard disk is recovered during the system recovery. For the
hard disk that had failed earlier and had been replaced (after purchase), its default setting
may not be the same as the factory state.
Strictly follow the steps described below. If you do any operation other
than the following steps, the system may not be recovered.
You need a keyboard for this operation.

Other manuals for Keysight Technologies E5061B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5061B and is the answer not in the manual?

Keysight Technologies E5061B Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5061B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals