Repair Guide 6
Keysight N8262A Service Guide 103
Use a T6 Torx screwdriver bit to remove the
three screws located on the bottom of the
chassis.
Use the 5/16” spanner to disconnect the power
reference input from the calibrator assembly.
Note:
You will be able to remove the calibrator
assembly at this step.
Instructions Visual