6 Repair Guide
116 Keysight N8262A Service Guide
Replacing the PPMC Assembly
The PPMC Assembly is pre-programmed with N8262A firmware.
Always perform a firmware update to the instrument if the PPMC assembly has
been replaced.
Instrument serial number:
This can be stored in the PPMC assembly via the command:
SERV:SNUM<CHARACTERDATA>
Instrument option(s):
This/these can be stored in the PPMC assembly via the command:
SERV:OPT“<CHARACTERDATA>”
Refer to the Programming Guide for further details on the use of these commands.