Theory of Operation 4
Keysight N8262A Service Guide 65
Mother Board Assembly
Purpose
Provides the average measurement path(s)
Provides the peak measurement path(s) to the measurement board(s)
Provides external trigger input/output and recorder output(s)
Provides signal routing between the PPMC and measurement board(s)
Inputs
Power supplies [from PSU]
Sensed power level(s) [from sensor flex(s)]
Trigger input [from external equipment]
Control, address, and data lines [from PPMC]
Outputs
Processed average path measurement [to PPMC]
Unprocessed peak path measurement samples [to measurement board(s)]
Trigger output and recorder output(s) [to external equipment]
Control and data lines [to PPMC]