Troubleshooting Guide 5
Keysight N8262A Service Guide 73
Power-Up Problems
Basic external checks
Check that the mains power source is live
Check the mains cable for any obvious damage
Basic internal checks
Check/reseat the cable between the line module and the PSU
Check/reseat the cable between the PSU and the mother board
Green LED DS1: If this is off, then the PSU may be faulty
Green LED DS4: This should come on when the power button is pressed
Green LEDs DS2/DS3: These will flash on and off during normal operation
Possible faults
–PSU
Mother board
Rocker switch defect
Loose front panel board ribbon cable