1 Specifications
38 Keysight N8262A Service Guide
Characteristics Peak Flatness
The peak flatness is the flatness of a peak-to-average ratio measurement for various
tone-separations for an equal magnitude two-tone RF input. Figure 1-2 refers to the
relative error in peak-to-average ratio measurements as the tone separation is varied.
The measurements were performed at –10 dBm with power sensors with 1.5 m cable
lengths.
Figure 1-2 N192XA error in peak-to-average measurements for a two-tone
input (High, Medium, Low, and Off filters)