1 Specifications
32 Keysight N8262A Service Guide
P-Series Wideband Power Sensor Specifications
The P-Series wideband power sensors are designed for use with the P-Series
power sensor only.
Maximum SWR
Sensor model Frequency range Dynamic range Damage level Connector type
N1921A 50 MHz to 18 GHz
–35 to +20 dBm
(500 MHz)
+23 dBm
(average power);
+30 dBm
(<1
s duration)
(peak power)
Type N (m)
–30 to +20 dBm
(50 to 500 MHz)
N1922A 50 MHz to 40 GHz
35 to +20 dBm
(500 MHz)
2.4mm (m)
–30 to +20 dBm
(50 to 500 MHz)
Frequency band N1921A/N1922A
50 MHz to 10 GHz 1.2
10 GHz to 18 GHz 1.26
18 GHz to 26.5 GHz 1.3
26.5 GHz to 40 GHz 1.5