Performance Tests 2
Keysight N8262A Service Guide 49
Zero Set (Average Path)
Description
Zero set is defined as the amount of residual offset error that is present following a
zero operation. This offset error is caused by contamination from several sources,
including circuit noise. This test measures the effectiveness of zero set by
performing 15 back-to-back zero operations of the average path (with no sensor
attached), after which the standard deviation of the results is calculated and
returned as the measured value.
Equipment
No test equipment required
Test Method
1 Execute the internal zero set measurement procedure for Channel A.
2 Read back the result of the measurement from the DUT.
3 Repeat this procedure for Channel B.
4 The test will take a few minutes to complete.
5 The measurement result should be less than0.0000175. The smaller the
measurement result, the smaller the amount of residual offset error.
NOTE
This test can be performed manually via the commands:
SERV:BIST:PEAK[1|2]:ZSET
SERV:BIST:CW[1|2]:ZSET:NUM?
(Refer to the Programming Guide for further details on the use of these
commands)
No adjustment is available for this test if it fails (see Chapter 5,
"Troubleshooting Guide").

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals