OUTPut:INHibit:MODE LATChing|LIVE|OFF
OUTPut:INHibit:MODE?
Sets the operating mode of the remote inhibit digital pin. The inhibit function shuts down the output in
response to an external signal on the Inhibit input pin. The Inhibit mode is stored in non-volatile
memory. See Programming the Digital Port.
LATChing - a logic-true signal on the Inhibit input causes the output state to latch OFF. The output
remains disabled until the Inhibit input is returned to logic-false and the latched INH status bit is
cleared by sending the OUTPut:PROTection:CLEar command or a protection clear command from the
front panel.
LIVE - allows the enabled output to follow the state of the Inhibit input. When the Inhibit input is true,
the output is disabled. When the Inhibit input is false, the output is re-enabled.
OFF - The Inhibit input is ignored.
Parameter Typical Return
LATChing|LIVE|OFF LATC, LIVE, or OFF
Sets the InhibitInput to Live mode: OUTP:INH:MODE LIVE
OUTPut:PON:STATe RST|RCL0
OUTPut:PON:STATe?
Sets the output power-on state. This determines whether the power-on state is set to the *RST state
(RST) or the state stored in memory location 0 (RCL0).Instrument states can be stored using the *SAV
command. This parameter is saved in non-volatile memory.
Parameter Typical Return
RST|RCL0 RST or RCL0
Sets the power-on state to the *RST state: OUTP:PON:STAT RST
l If the power-on state is set to 0 with no state stored, a self-test error "file not found; 0 state" is gen-
erated and the instrument is set to the *RST state.
l If a master auto-connect command fails, the power-on state is set to *RST.
OUTPut:PROTection:CLEar
Resets the latched protection. This clears the latched protection status that disables the output when
a protection condition occurs (see Programming Output Protection).
Parameter Typical Return
(none) (none)
Clears the latched protection status: OUTP:PROT:CLE
Keysight RP7900 Series Operating and Service Guide 213
5 SCPI Programming Reference