EasyManua.ls Logo

Keysight Technologies RP7945A

Keysight Technologies RP7945A
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
l Once started, you must complete each calibration section in its entirety. As each calibration sec-
tion is completed, the instrument calculates new calibration constants and begins using them.
However, these constants are not saved in nonvolatile memory until a SAVE command is explicitly
given.
l Exit the calibration mode either by logging out of the Admin menu or by sending CAL:STAT OFF.
Note that any calibration section that was calibrated but not saved will revert to its previous cal-
ibration constants.
Calibration Interval
One Year Calibration Interval
The instrument should be calibrated on a regular interval determined by the accuracy requirements of
your application. A one-year interval is adequate for most applications. Accuracy specifications are
warranted only if adjustment is made at regular calibration intervals. Published accuracy
specifications are not warranted beyond the one-year calibration interval.
Three Year Calibration Interval
Voltage and current programming and measurement accuracy specifications can be extended to a
three-year period calibration interval by multiplying (or increasing) the one-year calibration accuracy
specifications shown in the verification Test Record Forms by a factor of three.
Calibration Setups
6 Verification and Calibration
274 Keysight RP7900 Series Operating and Service Guide

Table of Contents

Related product manuals