Verification Setups
Test Considerations
For optimum performance, all verification and calibration procedures should comply with the
following:
l Ambient temperature is stable, between 18 and 28 °C.
l Ambient relative humidity is less than 80%.
l 30 minute warm-up period before verification or adjustment.
l Cables as short as possible, twisted or shielded to reduce noise.
6 Verification and Calibration
268 Keysight RP7900 Series Operating and Service Guide