EasyManua.ls Logo

Keysight Technologies RP7945A

Keysight Technologies RP7945A
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NPLC @60Hz Number of averaged points Voltage Measurement
% of range added to spec.
Current Measurement
% of range added to spec.
1 3255 0 0
0.6 1953 0.00013% 0.0005%
0.1 325 0.0005% 0.004%
0.06 195 0.0009% 0.005%
0.031 100 0.0015% 0.008%
0.010 33 0.0025% 0.0015%
0.006 20 0.004% 0.020%
0.003 10 0.006% 0.03%
0.0003 1 0.02% 0.10%
Excessive Dynamic Protection
For the high current models, EDP does not limit the allowable programmable waveform amplitude or
frequency. You can program a full scale square wave at a maximum frequency of 97.6 kHz indefinitely.
Output Quadrants
Model + V1 + V2 + V3 +/- I1 +/- I2 Minimum sink
resistance
RP7931A /RP7941A 20 V 12.5 V 0.5 V 400 A 250 A 1.25 mΩ
RP7933A /RP7943A 20 V 12.5 V 0.5 V 800 A 500 A 625 µΩ
RP7932A /RP7942A 80 V 40 V 0.75 V 125 A 62.5 A 6 mΩ
RP7935A /RP7945A 80 V 40 V 0.75 V 250 A 125 A 3 mΩ
RP7936A /RP7946A 160 V 80 V 1.5 V 125 A 62.5 A 12 mΩ
1 Quick Reference
50 Keysight RP7900 Series Operating and Service Guide

Table of Contents

Related product manuals