Keysight N9927-90001 User’s Guide 91
NA (Network Analyzer) Mode
NA Mode Settings
Multi-Trace Configurations
In NA Mode you can display multiple traces on the FieldFox screen.
The graphic above shows a x3H configuration. Tr2 is the ACTIVE trace as
indicated by the highlighted Tr2.
Trace Setting Notes
— The Frequency Range, IF BW, Resolution, Average, and Output Power
settings are common for all displayed traces.
— All other trace settings, such as measurement, format, and limit lines, are
applied individually to the ACTIVE trace in the same manner as when a
single trace is present.
— By default, a marker is created on ALL traces. However, they can be created
individually by disabling Coupled Markers. Learn more in
“Coupled Markers
(NA Mode)” on page 487
.
IMPORTANT! For CAT and NA modes, limit lines do not apply
where F1 = F2.