EasyManua.ls Logo

Keysight CX3322A - Connecting the CX1100

Keysight CX3322A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight CX3300 User’s Guide, Edition 4 21
Setting Up the CX3300
Installation
Connecting the CX1100
The CX1100 other than the CX1152A can be connected to the analog input
channels on the CX3300 front panel.
1. Attach the connector of the CX1100 to the desired channel.
2. Fasten the screws on the connector to assure the contact.
3. If a sensor head has not been connected to the CX1101A/CX1102A/CX1104A
current sensor yet, attach or connect it to t
he current sensor.
Figure 1-3 CX1100 Connection
CAUTION
Confirm the maximum input rating of the CX1100 and the sensor head. Do not
apply any electrical potential and/or current over the value.
NOTE
Connecting the CX1102A Current Sensor to the CX3324A
Connect the Primary and Secondary connectors to the contiguous channels
1 and 2, 2 and 3, or 3 and 4. Both connectors must be connected properly.
CX1100
Device under test
Sensor head
Analog input channels