Keysight CX3300 User’s Guide, Edition 4 97
Performing User Calibration
Using the CX1105A
Using the CX1105A
The CX3300 provides the offset user calibration for the CX1105A.
The user calibration data, Offset [V], is acquired and is used for error correction.
Effective for reducing the offset error.
Preparation: Before starting the user calibration, connect the CX1105A to the CX3300 analog
input channel.
Prepare the test adapter.
After the connec
tion, warm up them for 30 minutes.
Procedure: You can perform the user calibration as shown below.
1. Press the
Menu key several times to open the Configuration dialog box. The
Menu key is located above the Horizontal control on the front panel.
2. Click User Calibration to display the Configuration > User Calibration screen.
3. Click the Execute button for the channel under user calibration.
You will see
a dialog box which asks you to connect the test adapter to the
CX1105A input. Then connect them as shown in Figure 3-6 and click Yes.
The user calibration will start.