34 Keysight CX3300 User’s Guide, Edition 4
Setting Up the CX3300
Maintenance
Figure 1-9 CX1152A Troubleshooting
START
Replace the probe lead
of the failing channel with
a “working” probe lead.
END
Do you have
a “working” flat cable
for CX1152A?
Yes
No
Flat cable and/or pod
is defective.
Contact Keysight sales office
for further troubleshooting.
No
No
Replace the flat cable
with the “working” flat cable.
No
Pod is defective.
Contact Keysight sales office.
No
Replace the grabber
of the failing channel with
a “working” grabber.
No
Yes
Yes
Yes
Does the problem
still exist?
Does the problem
still exist?
Does the problem
still exist?