Keysight CX3300 User’s Guide, Edition 4 87
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Figure 3-1 Connection Diagram for User Calibration (CX1101A/CX1102A/CX1103A)
After the channel passed the user calibration, the ON/OFF switch will be 
available.
• Set ON to enable error correction.
Error correction is performed for the data measured by this channel from now 
on.
• Set OFF to disable error correction.
Error correction is not per
formed.
If the channel failed the user calibration, the ON/OFF switch is not available. 
Error correction is not performed.
Once you disconnect the CX1101A/CX1102A/CX1103A from the CX3300, the 
user calibration data for this channel will be cleared.
CX1101A/CX1102A with CX1203A
or 
CX1103A
SMA-to-BNC adapter 
(furnished with 
          current sensor)
BNC cable 
(furnished with CX3300)
CX3300