Keysight CX3300 User’s Guide, Edition 4 91
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Setting > Channels > Channel N screen:
Channel N: ON
Scale: 5 A/
Offset: 0 A
Bandwidth Limit: OFF
Invert Polarity: OFF
Setting > Sensor/Probe > Channel N screen:
Channel N: ON
Current Range: 10 A
Setting > Digital Channels screen:
Digital Channels: OFF
b. Control the current source (Keysight B2900) and apply the current of the
following co
ndition. For setting up the current source, refer to its User's
Guide.
Condition:
Pulse base: 0 A
Pulse peak: +10 A
Pulse width: 1 ms
Duty cycle: 2.5% for the Keysight B2900 series
c. Control the CX3300 and acquire the waveform as shown below.
1. Press the
Run/Stop key to start waveform acquisition.
2. Acquire the waveform as shown in Figure 3-3.
3. Press the
Run/Stop key to stop waveform acquisition.