208 Keysight Signal Generators Programming Guide
Programming the Status Register System 
Status Groups 
Data Questionable Calibration Event Enable Register
The Data Questionable Calibration Event Enable Register lets you choose which bits in the Data 
Questionable Calibration Event Register set the summary bit (bit 8 of the Data Questionable 
Condition register) to 1.
Command:
STATus:QUEStionable:CALibration:ENABle <value>, where <value> is the sum of the decimal 
values of the bits you want to enable. 
Example: Enable bit 1 and bit 0 so that whenever either of those bits are set to 1, the Data Questionable 
Calibration summary bit of the Data Questionable Condition Register is set to 1. Send the command 
STAT:QUES:CAL:ENAB 520 (2 + 1). 
Query: STATus:QUEStionable:CALibration:ENABle?
Response: Decimal value of the sum of the bits previously enabled with the 
STATus:QUEStionable:CALibration:ENABle <value> command.