Keysight Signal Generators Programming Guide 45
 Using IO Interfaces
Using LAN
2. At the command prompt, type in telnet.
3. Press the Enter key. The Telnet display screen will be displayed.
4. Click on the Connect menu then select Remote System. A connection form (Figure 2-5) is displayed. 
Figure 2-5 Connect Form (Keysight IO Library version J.01.0100)
5. Enter the hostname,  port number, and Te rm Type  then click Connect.
•Host Name−IP address or hostname
•Port−5023
•Term Type−vt100
6. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2-6 on page 46.
7. To signal device clear, press Ctrl–C on your keyboard.
8. Select  Exit from the Connect menu and type exit at the command prompt to end the Telnet 
session.
Using Telnet On a PC With a Host/Port Setting Menu GUI
1. On your PC, click Start > Run. 
2. Type  telnet then click the OK button. The Telnet connection screen will be displayed. 
3. Click on the Connect menu then select Remote System. A connection form is displayed. See 
Figure 2-5.
4. Enter the hostname,  port number, and Te rm Type  then click Connect.
•Host Name−signal generator’s IP address or hostname
•Port−5023
•Term Type−vt100
5. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2-6 on page 46.
6. To signal device clear, press Ctrl–C.
7. Select  Exit from the Connect menu to end the Telnet session.