58 Keysight Signal Generators Programming Guide
Using IO Interfaces
Using LAN
17. Disable LXI Output LAN Events on the MXA:
Send the following SCPI command:
:LXI:EVENt:OUTPut:LAN:DISable:ALL
18. Put the MXA into SA mode:
Send the following SCPI command:
:INST:SEL SA
19. Put the MXA into single sweep mode:
Send the following SCPI command:
:INIT:CONT OFF
20. Choose the MXA's PTP domain:
Send the following SCPI command:
:LXI:CLOCk:PTP:DOMain 0
The parameter value should match the one used in step 3.
21. Choose the MXA's LXI domain:
Send the following SCPI command:
:LXI:EVENt:DOMain 0
The parameter value should match the one used in step 4.
22. Move the MXA into the list sweep measurement:
Send the following SCPI command:
:CONF:LIST
23. Send the MXA a list of frequencies:
Send the following SCPI command:
:LIST:FREQ 100MHz,200MHz,300MHz,400MHz,500MHz
The MXA will put out signals at these frequencies and in this order.
24. Configure the MXA's list sweep parameters:
Send the following SCPI commands:
:LIST:ATT 10dB
:LIST:BAND:RES:TYPE FLAT
:LIST:BAND:RES 300kHz
:LIST:BAND:VID 3MHz
:LIST:SWE:TIME 1e-6 s
:LIST:TRIG:DEL 0
:LIST:DET RMS