Chapter 6: Replacing Assemblies
To remove and replace the internal digital input cable (MSO models only)
106
To remove and replace the internal digital input cable (MSO
models only)
Use this procedure to remove and replace the internal digital cable. When necessary,
refer to other removal procedures. The graphics in this chapter are representative of
the oscilloscope at the time of this printing. Your unit may look different.
1
Disconnect the power cable and remove the bottom sleeve.
2 Remove the AutoProbe board assembly from the front frame (see page 104).
3 Use a small screwdriver to remove the spring clip from the internal digital cable
connector then disconnect the cable from the acquisition board.
4 Clip the tie wraps holding the digital cable and shield to the side of the chassis.
5 Using a Torx T10 driver remove the screw holding the internal digital cable to
the front panel.
Figure 6-11
Removing the internal digital cable
6 Pull the internal digital cable out through the rear of the front panel.
7 To replace the internal digital cable, reverse the above procedure.
Shield
T10 torx
Internal
digital
cable
Acquisition
board
Tie wrap
Spring clip
Tie wrap