Setting Up the Oscilloscope 1
Keysight Infiniium S-Series Oscilloscopes User’s Guide 13
Environmental Characteristics
Environment Indoor use only
Ambient Temperature Operating: +5 °C to +40 °C
Non-operating: –40 °C to +65 °C
Humidity Operating: up to 80% relative humidity (non-condensing) at +40 °C
Non-operating: up to 90% relative humidity at +65 °C
Altitude Operating: up to 3,000 meters (9,842 feet)
Non-operating: up to 15,300 meters (50,000 feet)
Weight 26 lbs (11.8 kg)
Dimensions 16.8 in (43 cm) wide, 12.9 in (33 cm) tall, and 9 in (23 cm) deep
Safety ANSI/UL std. 61010-1:2012
CAN/CSA-22.2 No. 61010-1-12
Installation Category II
Voltage Fluctuations Note that the mains supply voltage fluctuations are not to exceed ±10% of the
nominal supply voltage.
Pollution Degree The Infiniium S-Series oscilloscopes may be operated in environments of Pollution
Degree 2.
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs. The
pollution has no influence. Example: A clean room or climate-controlled office
environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution
occurs which becomes conductive due to condensation which is expected.
Example: Sheltered outdoor environment.