Measurements 15
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 175
This measurement is available for analog channels.
Rising Edge Count
The Rising Edge Count measurement is an edge count for the selected waveform
source.
This measurement is available for analog channels.
Falling Edges Count
The Falling Edges Count measurement is an edge count for the selected waveform
source.
This measurement is available for analog channels.
Measurement Thresholds
Setting measurement thresholds defines the vertical levels where measurements
will be taken on an analog channel or math waveform.
Negative pulse count
NOTE
Changing default thresholds may change measurement results
The default lower, middle, and upper threshold values are 10%, 50%, and 90% of the value
between Top and Base. Changing these threshold definitions from the default values may
change the returned measurement results for Average, Delay, Duty Cycle, Fall Time,
Frequency, Overshoot, Period, Phase, Preshoot, Rise Time, +Width, and -Width.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight InfiniiVision 1200 X Series and is the answer not in the manual?

Keysight InfiniiVision 1200 X Series Specifications

General IconGeneral
BrandKeysight
ModelInfiniiVision 1200 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals