Utility Settings 22
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 241
1 Press [Utility] > Options > Auxiliary.
2 In the Auxiliary menu, press Gen Out; then, turn the Entry knob to select from:
• WaveGen— The waveform generator output. See Chapter 19, “Waveform
Generator,” starting on page 203.
• Triggers— The trigger output. Each time the oscilloscope triggers, a rising
edge occurs. The signal is is 0-5 V. See Chapter 11, “Triggers,” starting on
page 97.
• Mask— The mask test failure output. A 5 volt pulse occurs when the mask
test fails. When the mask test passes, no pulse occurs, and the signal
remains at 0 volts. See Chapter 16, “Mask Testing,” starting on page 179.
The output impedance of the Gen Out connector is 50 ohms.
Enabling Remote Command Logging
When remote command logging is enabled, remote commands sent to the
instrument (and results returned by the instrument) can be logged to the screen,
to a text file on a USB storage device, or to both the screen and a text file.
To enable remote command logging: