Reference 24
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 265
When using segmented memory, each segment is treated as a separate
waveform. All segments for a channel are saved, then all segments of the next
(higher numbered) channel are saved. This continues until all displayed channels
are saved.
When the oscilloscope is in the Peak Detect acquisition mode, the minimum and
maximum value waveform data points are saved to the file in separate waveform
buffers. The minimum value data points are saved first; then, the maximum value
data points are saved.
BIN data - using
segmented
memory
When saving all segments, each segment has its own waveform header (see
"Binary Header Format" on page 265).
In BIN file format, data are presented as follows:
• Channel 1 data (all segments)
• Channel 2 data (all segments)
• Channel 3 data (all segments)
• Channel 4 data (all segments)
• Digital channel data (all segments)
• Math waveform data (all segments)
When not saving all segments, the number of waveforms is equivalent to the
number of active channels (including math and digital channels, with up to seven
waveforms for each digital pod). When saving all segments, the number of
waveforms is equal to the number of active channels multiplied by the number of
segments acquired.
Binary Data in MATLAB
Binary data from an InfiniiVision oscilloscope can be imported to The MathWorks
MATLAB. You can download the appropriate MATLAB functions from the Keysight
Technologies web site at www.keysight.com/find/1200X-Series-examples.
Keysight provides the .m files, which need to be copied into the work directory for
MATLAB. The default work directory is C:\MATLAB7\work.
Binary Header Format
File Header There is only one file header in a binary file. The file header consists of the
following information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight InfiniiVision 1200 X Series and is the answer not in the manual?

Keysight InfiniiVision 1200 X Series Specifications

General IconGeneral
BrandKeysight
ModelInfiniiVision 1200 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals