N5186A MXG and X-Series SA Measurement Guide 15
Basic Measurements
Making Measurements
Using the equivalent SCPI commands
Creating a CW signal on Channel 1.
On the MXG:
SYSTem:PRESet
RF1:FREQuency:CW 1GHZ
RF1:POWer:AMPLitude 0dBm
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the Signal Analyzer:
FREQuency:CENTer 1GHZ
FREQuency:SPAN 1MHZ
CALCulate:MARKer1:MAXimum
On the MXG:
RF1:FREQuency:CW 1GHZ
RF1:POWer:AMPLitude -90dBm
RF1:OUTPut:STATe ON
On the Signal Analyzer:
DISPlay:WINDow1:TRACe:Y:RLEVel -70
DISPlay:WINDow1:TRACe:Y:PDIVision 5
BWIDth:VIDeo 300Hz
CALCulate:MARKer1:MAXimum
On the MXG:
RF1:FREQuency:CW 8GHZ
RF1:POWer:AMPLitude 0dBm
On the Signal Analyzer:
FREQuency:CENTer 8GHZ
DISPlay:WINDow1:TRACe:Y:RLEVel 0
DISPlay:WINDow1:TRACe:Y:PDIVision 10
BANDwidth:VIDeo:AUTO ON
CALCulate:MARKer1:MAXimum