56 N5186A MXG and X-Series SA Measurement Guide
Corrections
Corrections/De-embedding Using PathWave E7653APPC Software
Using the equivalent SCPI commands
Using a spectrum analyzer to make the corrections measurement
On the MXG:
SYSTem:PRESet
Set the power level to the highest level used in your measurement.
RF1:POWer:AMPLitude 5dBm
CORRection:PMDevice SANalyzer
CORRection:FLATness:STEP:STARt 5GHZ
CORRection:FLATness:STEP:STOP 7GHZ
CORRection:FLATness:STEP:POINts 20
CORRection:SANalyzer:COMMunicate:TYPE SOCKets
Set the LAN address and protocol parameters for your spectrum analyzer.
CORRection:SANalyzer:COMMunicate:LAN:IP "192.168.1.5"
CORRection:SANalyzer:COMMunicate:LAN:PORT 5025
CORRection:FLATness:CALibrate
CORRection ON
Using a power meter to make the corrections measurement.
On the MXG:
SYSTem:PRESet
RF1:POWer:AMPLitude 5dBm
CORRection:PMDevice PMETer
CORRection:FLATness:STEP:STARt 2.5GHZ
CORRection:FLATness:STEP:STOP 3.5GHZ
CORRection:FLATness:STEP:POINts 20
CORRection:SANalyzer:COMMunicate:TYPE USB
CORRection:PMETer:COMMunicate:USB:DEVice instr0
[Optional] CORRection:PMETer:CALibrate
[Optional] CORRection:PMETer:ZERO
CORRection:FLATness:CALibrate
Use query CORRection:PMETer:COMMunicate:USB:LIST? for a list of all
connected USB devices.