32 N5186A MXG and X-Series SA Measurement Guide
Basic Measurements
Making Measurements
Using the equivalent SCPI commands
On the MXG:
SYSTem:PRESet
RF1:FREQuency:CW 6GHZ
RF1:POWer:AMPLitude -10dBm
SIGNal1:MODE MTONes
SIGNal1:MTONe:ARB:NTON 15
SIGNal1:MTONe:ARB:FSP 15MHZ
SIGNal1 ON
RF1:OUTPut:STATe ON
For multi-channel instruments, set RF Out (all) to On.
RFALl:OUTPut ON
On the Signal Analyzer:
Change to Spectrum Analyzer mode, Swept SA measurement.
INSTrument:CONFigure:SA:SAN
DISPlay:VIEW:ADVanced:SELect “NORMAL
FREQuency:CENTer 6GHZ
FREQuency:SPAN 250MHZ
BANDwidth:RESolution 30KHZ