3-26 Service Guide N5227-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
System Verification N5227A
Interpreting the Verification Results
Figure 3-11 shows an example of typical verification results with Print Tabular Data selected in the Printer
Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of the device, and the date
tested.
Each S-parameter measurement result is printed with frequency tested, lower and upper limit lines, the
measured data, and the result of the test.
Figure 3-11 Example of Printed Tabular Verification Results