[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the current, voltage, power, or resistance before and after the pulse occurs for Arb sequence
steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<start value>
Step 0 - 99, *RST 0 (none)
Programs a start- and end-level current: ARB:CURR:PULS:STAR 1,(@1)
Programs a start- and end-level voltage: ARB:VOLT:PULS:STAR 1,(@1)
Programs a start- and end-level power: ARB:POW:PULS:STAR 1,(@1)
Programs a start- and end-level resistance: ARB:RESISTANCE:PULS:STAR 1,(@1)
Programs a start- and end-level for step 0: ARB:SEQ:STEP:CURR:PULS:STAR 1, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:STARt:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:STARt:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the pulse occurs for Arb
sequence steps.
5 SCPI Programming Reference
196 Keysight N6700C Operating and Service Guide