Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 0 <start time>
Step 0 - 99, *RST 0 (none)
Programs a start time for the current pulse: ARB:CURR:PULS:STAR:TIM 1,(@1)
Programs a start time for the voltage pulse: ARB:VOLT:PULS:STAR:TIM 1,(@1)
Programs a start time for the power pulse: ARB:POW:PULS:STAR:TIM 1,(@1)
Programs a start time for the resistance pulse: ARB:RESISTANCE:PULS:STAR:TIM 1,(@1)
Programs a start time for step 0: ARB:SEQ:STEP:CURR:PULS:STAR:TIM 1, 0,(@1)
[SOURce:]ARB:CURRent:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:TOP[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:TOP[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the top current, voltage, power, or resistance level of the pulse.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:TOP[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:TOP[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the top current, voltage, power, or resistance level of the pulse for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<top value>
Step 0 - 99, *RST 0 (none)
Programs the top level current: ARB:CURR:PULS:TOP 2,(@1)
Programs the top level voltage: ARB:VOLT:PULS:TOP 20,(@1)
Programs the top level power: ARB:POW:PULS:TOP 50,(@1)
Programs the top level resistance: ARB:RESISTANCE:PULS:TOP 20,(@1)
Programs the top level for step 0: ARB:SEQ:STEP:CURR:PULS:TOP 2, 0,(@1)
Keysight N6700C Operating and Service Guide 197
5 SCPI Programming Reference