Step
Back to ARB Commands
[SOURce:]ARB:CURRent:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the step current, voltage, power, or resistance level.
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the step current, voltage, power, or resistance level for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<end value>
Step 0 - 99, *RST 0 (none)
Programs the step level current: ARB:CURR:STEP:END 2,(@1)
Programs the step level voltage: ARB:VOLT:STEP:END 20,(@1)
Programs the step level power: ARB:POW:STEP:END 50,(@1)
Programs the step level resistance: ARB:RESISTANCE:STEP:END 100,(@1)
Programs a step level for step 0: ARB:SEQ:STEP:CURR:STEP:END 2, 0,(@1)
Keysight N6700C Operating and Service Guide 211
5 SCPI Programming Reference