[SOURce:]ARB:CURRent:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:POWer:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the initial current, voltage, power, or resistance level.
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STEP:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:STEP:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STEP:STARt[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:STEP:STARt[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the initial current, voltage, power, or resistance level for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<start value>
Step 0 - 99, *RST 0 (none)
Programs a start level current: ARB:CURR:STEP:STAR 1,(@1)
Programs a start level voltage: ARB:VOLT:STEP:STAR 1,(@1)
Programs a start level power: ARB:POW:STEP:STAR 1,(@1)
Programs a start level resistance: ARB:RESISTANCE:STEP:STAR 50,(@1)
Programs a start level for step 0: ARB:SEQ:STEP:CURR:STEP:STAR 1, 0,(@1)
[SOURce:]ARB:CURRent:STEP:STARt:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STEP:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:STEP:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:STEP:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:STARt:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:STEP:STARt:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the delay in seconds after the trigger is received, but before the step occurs.