SENSe:SWEep:POINts <value>, (@<chanlist>)
SENSe:SWEep:POINts? [MIN|MAX,] (@<chanlist>)
Defines the number of points in a measurement. The number of points depends on the line frequency
(50 Hz or 60 Hz). The default number of points result in a measurement of 1 NPLC.
Parameter Typical Return
1 to 524,288|MIN|MAX,
*RST 3255 (60 Hz); 3906 (50 Hz)
<points>
Specifies 2048 points: SENS:SWE:POIN 2048, (@1)
SENSe:SWEep:TINTerval <value>, (@<chanlist>)
SENSe:SWEep:TINTerval? [MIN|MAX,] (@<chanlist>)
Defines the time period between measurement samples. Units are in seconds. Values are rounded to
the nearest 20.48 microsecond increment. Below 20.48 microseconds, values are rounded to the
nearest 10.24 or 5.12 microsecond increment respectively.
Parameter Typical Return
0.00000512 to 40,000|MIN|MAX, *RST 0.00000512 <time interval>
Specifies an interval of 1 ms between points: SENS:SWE:TINT0.001, (@1)
SENSe:SWEep:TINTerval:RESolution RES20|RES40
SENSe:SWEep:TINTerval:RESolution?
Sets the resolution used in setting the time interval between digitized measurements. Applies for time
intervals above 20.48 microseconds. External data log measurements also use this setting.
RES20 - rounds all values above 20.48 microseconds to the nearest 20.48 µs multiple.
RES40 - rounds all values above 20.48 microseconds to the nearest 40.96 µs multiple.
Parameter Typical Return
RES20|RES40, *RST 20.48 µs RES20 or RES40
Specifies a 40 µs resolution: SENS:SWE TINT:RES RES40
SENSe:VOLTage[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:VOLTage[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects a DC voltage measurement range. The value that you enter must be the highest value that
you expect to measure. Units are in volts. The instrument selects the range with the best resolution
for the value entered.