3 VMA Mode
3.11 Phase and Amplitude vs Time Measurement
Preset 180
State Saved Saved in instrument state
Min 1
Max 1000
Total Meas Time
Information only. Returns the result of (Meas Offset + Meas Interval + Meas
Transient)*Segments.
Remote Command
[:SENSe]:PAVTime:MTIMe?
Example
:PAVT:MTIM?
State Saved Saved in instrument state
Meas Interval
Sets the duration of acquisition window within the segment.
Remote
Command
[:SENSe]:PAVTime:SEGMents:INTerval <time>
[:SENSe]:PAVTime:SEGMents:INTerval?
Example
:PAVT:SEGM:INT 10 us
:PAVT:SEGM:INT?
Couplings Meas Interval must be > 0 s and less than (Max Capture Length/Segments) - Meas Offset - Meas
Transient
Preset 1 ms
State Saved Saved in instrument state
Min 100 ns
Max (Max Capture Length/Segments) - Meas Offset - Meas Transient
Meas Offset
Sets the offset prior to "Meas Interval" on page 1822 within the segment.
Remote
Command
[:SENSe]:PAVTime:SEGMents:OFFSet <time>
[:SENSe]:PAVTime:SEGMents:OFFSet?
Example
:PAVT:SEGM:OFFS 1 us
:PAVT:SEGM:OFFS?
Couplings Meas Offset must be >= 0 s and less than (Max Capture Length/Segments) - Meas Interval - Meas
Transient