6 Input/Output
6.1 RF Source
Step Configuration of Step Duration parameter list (Remote Command Only)
Configures the Step Duration parameter array of the whole List Sequencer at one
time. The number of arrays is same as step number defined in "Number of Steps" on
page 2028. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:DURation:TYPE <enum>, <enum>, <enum>,
:SOURce:LIST:SETup:DURation:TYPE?
Example
:SOUR:LIST:SET:DUR:TYPE COUN,TIME,CONT
:SOUR:LIST:SET:DUR:TYPE?
Notes Sets up or queries the parameter array of whole list sequence
Step Duration <enum> - specifies the duration of the step. The duration can be specified to be either
time, or play count of the ARB file associated with the step, or continuous. If Waveform is set to “CW”,
this value cannot be set to Play Count and an error will be generated. If continuous is selected, the
following Time or Count value is ignored. For further details of this setting, see "Step Duration" on
page 2038
Options are:
TIME | COUNt | CONTinuous
If input parameter number exceeds the step number defined by "Number of Steps" on page 2028,
then error -221, "Settings conflict; The number of input parameters is too large and is truncated to
current list step number" is generated, and only those parameters whose index number falls within
number of steps will be updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 2028
Step Configuration of Duration Time or Play Count parameter list (Remote Command
Only)
Configures the Duration Time or Play Count parameter array of the whole List
Sequencer at one time. The number of arrays is same as step number defined in
"Number of Steps" on page 2028. As a step is set up, the value entered runs through
several levels of validation.
Remote Command
:SOURce:LIST:SETup:TOCount <time/int>, <time/int>, <time/int>,
:SOURce:LIST:SETup:TOCount?
Example
:SOUR:LIST:SET:TOC 1s,2s,3s
:SOUR:LIST:SET:TOC?
:SOUR:LIST:SET:TOC 5,6,7
:SOUR:LIST:SET:TOC?
Notes Sets up or queries the parameter array of whole list sequence