4 System
4.6 Alignments
Only functional for the following alignments:
Align Now All of VXT models M9410A/11A and M9410E/11E
Align Source
Align Receiver
Align Fast
Align LO Leakage
Align VXT Transceiver of M910E/11E
Align Up Down Converter of M9410E/11E
Align Selected Freq Ranges only guarantees the hardware performance within the frequency range
Preset
OFF
Enable Extended Freq Range
Allows you to set frequency ranges for VXT models M9410A/11A/15A with Remote
Head and/or CIU. When Enable Extended Freq Range is not active, the frequency
range is limited by VXT models only.
Remote Command
:CALibration:INTernal:ASFRanges:EXTend[:STATe] ON | OFF | 1 | 0
:CALibration:INTernal:ASFRanges:EXTend[:STATe]?
Example
:CAL:INT:ASFR:EXT ON
:CAL:INT:ASFR:EXT?
Dependencies Only available in VXT models M9410A/11A/15A/16A with Remote Head and/or CIU
Only available when Align Specified Freq Ranges is ON
Preset
OFF
Frequency Range
Allows you to set the alignment frequency range.
Remote Command
:CALibration:INTernal:ASFRanges:FRANges <startFreq>,<stopFreq>[,<startFre-
q>,<stopFreq>][,<startFreq>,<stopFreq>][,<startFreq>,<stopFreq>][,<startFre-
q>,<stopFreq>]
Example
:CAL:INT:ASFR:FRAN 1.3 GHz,1.8 GHz,2.5 GHz,3.9 GHz
:CAL:INT:ASFR:FRAN?
Notes <startFreq>: Start frequency of an alignment
<stopFreq>: Stop frequency of an alignment
To process alignment for a single frequency point, set <startFreq> = <stopFreq>
1917 Vector Modulation Analyzer Mode User's &Programmer's Reference