EasyManua.ls Logo

Keysight N9041B - Meas Preset; 3.4.8.2 Advanced; Phase Noise Optimization

Keysight N9041B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.4 IQ Waveform Measurement
Dwell Time
Range Table Resolution Bandwidths
Range Table Dwell Times
Meas Preset
Restores all measurement parameters to their default values.
Remote Command
:CONFigure:WAVeform
Example
:CONF:WAV
Notes Restore default values of all parameters
3.4.8.2 Advanced
Contains advanced functions that are used for specific applications. These settings
should not be changed for most measurements.
Does not appear in VXT.
Phase Noise Optimization
Sets the LO (local oscillator) phase noise behavior for various desired operating
conditions.
Remote Command
[:SENSe]:WAVeform:FREQuency:SYNThesis[:STATe] 1 | | 5
For the meaning of each numeric option value, see "Parameter Options & Installed Options" on page
768 below
[:SENSe]:WAVeform:FREQuency:SYNThesis[:STATe]?
Example
:WAV:FREQ:SYNT 2
Selects optimization for best wide offset phase noise
Dependencies Does not appear in all models. For models that do not display this control, the SCPI command is
accepted for compatibility, but no action is taken
Not available in VXT models M9410A/11A/15A/16A and M9410E/11E/15E/116E
Preset Because this function is in AUTO after Preset, the state of this function after Preset will be automat-
ically calculated
State Saved Saved in instrument state
Range See "Ranges" on page 772 below
Auto Function
767 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents