7 Save/Recall/Print
7.3 Save
Only one Transducer units can be on at any given time. Note that this means that if a
correction file with a Transducer Unit is loaded into a particular Correction, all other
Corrections are set to that same Transducer unit. Note that the legacy term
“Antenna Unit” is still used in the correction file, even though the more modern term
“Transducer Unit” is used in the user interface.
Similarly, the Bias rows can only be used in Correction register 1, because there can
only be one setting for Bias at any given time. If a Correction file with a Bias or Bias
State row is loaded into any Correction register but 1, an error is generated: Mass
storage error; Can only load Bias Settings into Correction 1
The data follows the DATA row, as comma separated X, Y pairs; one pair per line.
For example, suppose you have an Antenna to correct for on an N9020A version
A.02.06 and the correction data is:
0dB at 200MHz
17dB at 210MHz
14.8dB at 225MHz
Then the file will look like:
Amplitude Correction
"Correction Factors for 11966E"
"Class B Radiated"
A.02.06,N9020A
P13 EA3 UK6,01
Frequency Unit, MHz
Antenna Unit, dBuV/m
Frequency Interpolation, Linear
DATA
200.000000,0.00
210.000000,17.00
225.000000,14.80
The choices for the 1 of N fields in the metadata are as follows:
2406 Vector Modulation Analyzer Mode User's &Programmer's Reference