69
Functions and Measurements
Making an Average Power Measurement
To Make a Basic Average Power Measurement
Follow the steps below to make a basic average power
measurement.
1 Press > Preset. (With Preset Type of Factory)
2 Press , rotate the knob or press the arrow hardkey to
highlight Power Meter, then press [ENTER] to turn on the
power meter.
3 Zero and calibrate the meter. Press {Zero} to make a Zero
operation of the power sensor followed by a calibration
operation.
4 Connect the power sensor to the power Ref 50 MHz port.
The analyzer supports the U2000 Series power sensors.
5 Connect the external attenuation, if required, with the power
sensor used.
6 Connect the analyzer to the base station. Connect the power
sensor to the signal to be measured. A typical BTS
connection is after the output of the power amplifier and
duplexer. The connection requires a 2 ft. cable and two
Type- N barrel connectors, in addition to the power sensor
and external attenuator.
Figure 3-25 Connection with base station
Preset/
System
MODE