EasyManua.ls Logo

Keysight u2722a - Page 59

Keysight u2722a
71 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation and Features 2
Keysight U2722A/U2723A User’s Guide 59
SOUR:MEM:LIST 1, (@1) This command selects memory list 1
as the active memory list for channel
1. This command can be omitted if the
default active memory list has not
been changed.
SOUR:MEM:LIST:CLE, (@1) This command clears all the content in
the active memory list of channel 1.
SOUR:MEM:VOLT:RANG R20V, (@1) This command stores the “set voltage
range to 20 V” command into the
active memory list of channel 1.
SOUR:MEM:CURR:RANG R120mA,
(@1)
This command stores the” set current
range to 120 mA” command into the
active memory list of channel 1.
SOUR:MEM:CURR:LIM 0.12, (@1) This command stores the “set current
limit to 120 mA” command into the
active memory list of channel 1.
SOUR:MEM:VOLT:SOUR 20, (@1) This command stores the “drive
voltage 20 V” command into the active
memory list of channel 1.
SOUR:MEM:OUTP ON, (@1) This command stores the “set output
on” command into the active memory
list of channel 1.
SOUR:MEM:CURR:MEAS, (@1) This command stores the “measure
current” command into the active
memory list of channel 1.
SOUR:MEM:LIST:STOR, (@1) This command stores all commands
from the active memory list of channel
1 into the nonvolatile memory. This
command can be omitted if this
operation is not required.
SOUR:TRIG:STRG This command configures the
backplane trigger to star trigger.
SOUR:MEM:ARM (@1) This command sets the arm type of
channel 1 to the memory list arm.
INIT:TRAN, (@1) This command arms channel 1.