2 Operation and Features
60 Keysight U2722A/U2723A User’s Guide
Execute commands from memory list using the remote trigger command
Commands from the memory list can be executed by sending the remote trigger 
command. There will be some small time delay between the channels if two or 
three channels are triggered by the remote command at the same time. The 
channel execution sequence will start in the order of channel 1, channel 2, and 
channel 3.
Example 2
This example stores the source current and measure voltage commands into the 
active memory list (memory list 1) for channel 1 and executes them when a 
remote trigger command is received.
To perform the function above, follow the sequence below.
a Select an active memory list.
b Clear the content of the active memory list.
c Store the “set voltage range” command into the active memory list.
d Store the “set current range” command into the active memory list.
e Store the “set voltage limit” command into the active memory list.
f Store the “drive current” command into the active memory list.
g Store the “set output on” command into the active memory list.
h Store the “measure voltage” command into the active memory list.
i Trigger the execution of commands in the active memory list.
→ SOUR:MEM:LIST 1, (@1) This command selects memory list 1 
as the active memory list for channel 
1. This command can be omitted if 
the default active memory list has not 
been changed.
→ SOUR:MEM:LIST:CLE, (@1) This command clears all the content 
in the active memory list of channel 1.
→ SOUR:MEM:VOLT:RANG R20V, (@1) This command stores the “set voltage 
range to 20 V” command into the 
active memory list of channel 1.