6 Read the stored files
After a controlled stopping of the measurement and a data storage, it's possible to see on the screen measurement dataset results.
6.1 Access to files
Press Menu key then select Read then OK.
The first stored file is displayed.
Scroll stored files with the scrolling knob key.
Each file is identified by a serial number (type C01) and a filename (format : S01_1811LPF). This file name will appear when uploading
to a computer.
For easy identification, the shown screen includes all measurement configuration :
Mode : L-St, L-Leq, Leq-St, S1 + S2.
According to the mode : different frequency and time weightings, logging time, sampling rate, date and time...
6.2 Read all data file
6.2.1 L-St mode file
When the file has been identified by its identification number:
Press OK key to access to E1 main screen results.
LUpk : level of maximum peak of sound pressure
LXYmax : maximum value of sound pressure level weighted temporally met since the beginning of the measurement.
LXYmin : minimum value of sound pressure level weighted temporally met since the beginning of the measurement.
D : Measurement time expressed in DD / HH: MM: SS.
% : percentage of saturation of the input stage.
Statistics indices : L01 -L10 – L50- L90- L95
With scrolling knob key, go to E2 screen which reminds measurement configuration.
Date of measurement start
Date of measurement end
Measurement time expressed in DD / HH: MM: SS.
Measurement mode : L-St
Frequency weighting : Weig.
Time weighting : Cst
Frequency weighting of the level of peak pressure : LpK
Sampling rate : Samp.
21 Read the stored files