6.2.2 L-Leq mode file
When the file has been identified by its identification number:
Press OK key to access to E1 main screen results.
LUpk : level of maximum peak of sound pressure
LXYmax : maximum value of sound pressure level weighted temporally met since the beginning of the measurement.
LXYmin : minimum value of sound pressure level weighted temporally met since the beginning of the measurement.
LXeq : equivalent continuous level of acoustic pressure frequency weighted on measurement time.
LXE : level of exposure to sound frequency weighted
D : measurement time expressed in DD / HH: MM: SS.
% : percentage of saturation of the input stage.
With scrolling knob key, go to E2 screen which reminds measurement configuration.
Date of measurement start
Date of measurement end
Measurement time expressed in DD / HH: MM: SS.
Measurement mode : L-Leq
Frequency weighting for L and Leq : Weig.
Time weighting for L : Cst
Integration time for leq : S/S (Start/Stop mode)
Frequency weighting of the level of peak pressure for L : LpK
Press Esc key to quit this file and back to file selection main screen.
6.2.3 Leq-St mode file
When the file has been identified by its identification number:
Read the stored files 22
Stored file
E1 screen E2 screen Stored file
Stored file
E1 screen E2 screen Stored file

Table of Contents