1919
GB
Use of the Use of the Kinetec Patient Pad KitKinetec Patient Pad Kit
The KINETEC Patient Pad Kit is designed for rapid fitting, optimal hygiene and maximum patient comfort.
- For using and positioning the straps, please refer to here under. Make sure that the self-adhesive parts (26) are visible.
- Place the sponge side next to the skin.
FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
Each cover is provided with a label to record the patient’s nameEach cover is provided with a label to record the patient’s name
CLEANING:
- Sterilization of the pads ( if necessary): Sterilized at 134 °C during 18 minutes.
- Disinfecting of the pads: Washing at 30°C with use of a disinfecting solution during the rising cycle.
Example of product that can be used: Solution " Baclinge" at 0.125 % or "Souplanios" at 0,125% from ANIOS Laboratory.
A complete list of distributors in your country is available on request.
The KINETEC Spectra is delivered with a complete set. Components:
- 4 straps
- 1 foot support
- 1 cover