2323
GB
Product informationProduct information
CleaningCleaning
Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
Technical specificationsTechnical specifications
Product:Product:
Weight: 11Kg (24 pounds)
Splint dimensions: 94cm (37 inches) x 33cm (13 inches) x 33cm (13 inches)
Angular limits: -10° to 120°
Speed: from 30 to 160° per minute for average femur length
Patient height: full leg: 71 to 104 cm (28 to 41 inches)
Tibia: 38 to 58 cm (15 to 23 inches)
Femur: 33 to 46 cm (13 to 18 inches)
Electricity:Electricity:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm (KINETEC order: 4610007434)
EnvironmentEnvironment
- Storage/transport conditions: Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
- Operating conditions: Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%