The standard gateway of each unit needs its own
IP address that is not yet in use in your local net-
work.
At the factory, the IP address of the standard
gateway is set to: 192.168.0.200
1. Insert unit key in key switch.
2. Set unit key to “P”.
3. Press and hold [Max.] and [Min.] simultaneously for four sec-
onds.
ð
The display shows .
4.
Press [Max.] and [Min.] to go to level and confirm with
Set.
5.
Press [Min.] to select the parameter .
6. Press and hold [Target temperature].
ð
The first three digits of the IP address are displayed.
7. Use [Max.] and [Min.] to set the selected digits of the required
IP address.
ð
The number changes on the display in steps of one.
8. Release [Target temperature].
ð
The first three digits of the IP address have been
entered.
9.
Press [Min.] to select the next parameter .
10. Press and hold [Target temperature].
ð
The next digits of the IP address are displayed.
11. At levels L62 and L63 repeat steps 6 to 10 until the IP
address of the standard gateway has been entered com-
pletely.
12. Set key switch to “1” to complete your input.
ð
The IP address of the standard gateway has been set up.
Setting up the IP address of the
standard gateway on the unit
Operation
Setting up PC-KIT-NET (optional)
27.05.2021 Refrigerator/freezer LABEX® series 47