Personnel:
n
System/network administrator (recom-
mended)
Materials:
n
Network cable (EIA/TIA-568 standard),
(included in the scope of delivery of PC-
KIT-NET)
Requirements:
n The TCP/IP module has been installed (
Ä
Chapter 1.7 “Addi-
tional unit components” on page 18 and
“Assembly and con-
nection manual – PC-KIT-STICK/PC-KIT-NET”).
n The KIRSCH-DATANET software (version 5.0 or higher) has
been installed on the local PC or local network.
1.
Switch off the unit
Ä
Chapter 7.2 “Switching off the unit”
on page 42.
2. Unplug the power plug.
3. Remove the dust cover from TCP/IP output.
4. Insert the network cable in the TCP/IP output.
5. Connect the network cable to the network socket.
ð
The unit has been connected to the network.
6. Insert the power plug in the socket.
7.
Switch on the unit
Ä
Chapter 7.1 “Switching on the unit”
on page 42.
1. Insert unit key in key switch.
2. Set unit key to “P”.
3. Press and hold [Max.] and [Min.] simultaneously for four sec-
onds.
ð
The display shows .
4.
Press [Max.] and [Min.] to go to level and confirm with
Set.
5.
Press [Min.] to select the parameter .
ð
The first digits of the MAC address are displayed.
Connecting the unit to the local
network
Reading out the MAC IP address on
the unit
Operation
Setting up PC-KIT-NET (optional)
27.05.2021Refrigerator/freezer LABEX® series48